Friday, July 10, 2009

Not Much Happening...

Today (Wednesday) was a bit slower than the past few days. I spent the majority of my time at lab catching up with note-taking and playing around with MATLAB a little bit. Aside from this, I plan on taking at least one more lateral scan across the wafer to try and characterize the THz beam. I hope to get this done tomorrow.

Thursday, July 9, 2009

More Lateral Scans with a New Lens

I spent most of the day today (Tuesday) looking at the Fabry-Perot effect which is easy to see in the temporal pulses. Take the following plot:



This is a plot of the (0,0) position on the Si wafer. Like I said in a recent post, there are three distinct pulses -- the initial pulse which has the greatest amplitude which is then followed by a pulse of slightly less amplitude and then yet another peak of slightly less amplitude. If we look at the horizontal scale, which is measured in picoseconds, we can see that the difference between peaks is approximately 12 ps. We also notice that the three peaks are equally spaced with this same increment.

The math for this calculation is pretty straight-forward. Basically, we have the Si wafer of width l and refractive index n. Most of the signal transmits directly through the wafer, and is seen in the primary pulse in the above scan. Then some of the signal reflects off the boundary between the back of the Si disc (call it b), then reflects onto the first boundary which it has already passed through between the front of the Si and the air (call this a), and finally back through b. This gives us that second pulse. The third pulse is explained in precisely the same fashion -- signal transmits through a, reflects off b, reflects off a, reflects off b, reflects off a, and is then transmitted through b. Since the thickness of the wafer determines the path length of each piece of the pulse, they each show up at a different position on the time scale.

I understand the basic physics behind this effect, but I am interested in determining a method to effectively remove it from the data.

I also spent some time in changing the lens used to focus the excitation beam. We found a lens which had a longer focal length so as to get a tighter focus of the excitation beam on the Si wafer.

After doing this, I took more lateral scans across the Si wafer to determine if the beam shape is still about the same as before. The data that I found are shown in the plots below. First is that of the x- and y-direction on the same set of axis.


The scan traced with the blue line and markers is of the x-direction, while the red line and markers display the scan for the y-direction. We can see from the above plot that the beam profile for the x-direction seems to be even more distorted than what we saw when using the other lens. The reason for this, as Antoine explained, might be due to a low-pass filter effect in which the size of the excitation beam effectively changes the profile of the THz beam. Basically, since the excitation beam is so much tighter and we are creating more accurate measurements than with the wider beam from before, we may expect some of this sort of over-distortion to come about. We also see the same sort of skewness in this scan as the x-scan that was done with the other lens.

The y-scan shows the same sort of Gaussian distribution from before. It is still rather symmetric and looks relatively normal. It is also interesting to note that the peak amplitudes for the two scans do not match up perfectly. I am not really sure what this is due to, since I took both scans within about 30 - 60 minutes of each other.

The distortion in the x-direction and the normal, Gaussian shape in the y-direction suggest that perhaps the x-incidence excitation beam is distorted in some way and thus we see such strange profiles for the x-scans.

To further investigate the matter, I took another scan in the x-direction but at about -HM of the y-direction. This plot is shown below.


It is clear that we see the same sort of distortion in this scan... and skewness. Again, I feel as though the reason for this might be the incidence of the x-beam on the wafer. It would be interesting from here to take even more lateral scans and try to actually build a beam profile from this. I might look into taking what one might consider to be circular contours about the xy-plane to see if I can create a better 3D image, though there may be more interesting things to do.

Tuesday, July 7, 2009

Spectral Ratios and More

I spent the day (Monday) looking at the scans I took last Friday. I imported the data and made a variety of different plots. The main things that I looked at were:

Temporal signal with spectrum side-by-side
Spectra of all the data on the same set of axes
Spectra of positive and negative HM of both the x- and y-directions
Ratio of two spectra

I found some interesting things, but did not spend too much time on quantifying the data. Following are a series of images from each of the four characteristic plots that I mention above.

First is a plot which shows the temporal pulse and the spectrum of this pulse in side-by-side plot windows.

This happens to be the average over four sets of data at the maximum, or most intense, part of the THz beam. In the temporal pulse to the left, it is noticeable that there is the initial peak of the pulse and then shortly thereafter and spaced equally apart there are two other peaks. We think this is due to a Fabry-Perot effect in which the THz signal reflects within the Si before transmitting. I would like to write a script which removes this effect, as it has an influence on the spectrum.

The plot to the right shows the spectrum of this same temporal pulse. The green is the actual spectrum while the blue is the same spectrum after being smoothed out.

Following is a plot which shows the spectra of at each point. This includes the maximum position and the positions for the HM of both directions.

The legend shows the coordinate for each of these spectral scans. It is interesting that the +/- scans at the FWHM in the x-direction do not match up better. This is a little bit more clear in the following plot, which compares the +HM to the -HM of the x-direction.

There is clearly some great variation in these two spectra. Recall that the beam did appear to be skewed in the x-direction, and perhaps this is part of the explanation. We see some different things in the y-direction, which, contrary to the x-direction, seemed to be more symmetric. This plot is shown below.

These spectra seem to be much more symmetric, though there is that sharp valley for the spectra of the +HM direction (recall that I have labelled it ++HM since the other value of +HM seemed to be off). I have not yet taken the time to try and explain this sharp drop.

In order to look at the spectra relative to the reference spectra, I divided the spectra at each position into the spectra for (0,0). A sample plot of this is shown below.

This happens to be the ratio of the (+HM,0) position. This is plotted on a double y-axis plot to show the ratio in green (corresponding to the right vertical axis) as well as the spectrum of the given position (in red) and the spectrum of the maximum position (in blue). Both of the spectra correspond to the left vertical axis.

The idea of taking this ratio is to determine the beam waist size as a function of frequency, though I need to read into this more.

A few more things to look at are possibly taking an even better spatial profile of the beam since there is such a difference in the x-position. This will also allow me to find a better estimate of the FWHM in both x- and y-directions. I also need to look at how our excitation beam may be destroying the Si and thus determine if this effect is negligible for our purposes or not. Also, there is the idea of removing most of the moisture content of the air via replacing it with gaseous nitrogen.

Monday, July 6, 2009

Changing the Time Constant and Taking More Scans

As I mentioned yesterday, I wanted to take some more precise scans today (Friday) and also learn how to plot the ratio of the two scans via MATLAB. I was able to take data at all five postions mentioned earlier:

(0,0)
(+HM,0)
(-HM,0)
(0,+HM)
(0,-HM)

I changed the time constant from 1 second per point (as we used earlier this week) to 100 ms. This changed the length of the scan to take about 9 min instead of the 90 min from earlier in the week. I took four scans at each of the above positions using the positions of FWHM that Antoine had determined. The only problem that I ran into was the position for the (0,+HM) gave scans a lot different from the others and the position was about 0.5 mm different from what I estimated on my initial data. Because of this, I took one more scan at the point I had expected the FWHM to be at. I labelled this point as:

(0,++HM)

For the first five positions listed above, I took data in a cyclic permutation, but for the point (0,++HM), I just took the four scans all right after each other.

During these scans, I was able to update my lab notebook and I made a note to look at how a Fabry-Perot effect might alter this data.

I tried to learn how to take the ratio with MATLAB but was unsuccessful. I talked to Guilhem about this and he gave me some commands to use. I plan on trying this out either this weekend or sometime next week. I should also have plots of the data early next week.

Spectral Plots and More MATLAB Functions

The plan for today (Thursday) was to learn more with MATLAB and to take some more scans at the FWHM of each x- and y- directions as well as the maximum. This is different from what we did earlier in the week, as I plan to take data at all five positions (max and FWHM for each) and take a series of shorter scans instead of one long scan. Then I can average the data for all of these scans, which should take better care of possible anomolies.

Unfortunately, I did not have time to take more data today. Instead I worked with MATLAB and importing data from the scans. To do this, I used specific functions that Guilhem wrote to handle such data. With said functions, it is possible to import and build the signal we initially took (temporal). Given a series of indexed data like what I mentioned above, this function will average those signals into one. The other important function was to take this temporal signal and create the spectrum.

I was able to make a few plots with this, one of which shows the spectra of each scan. This plot is shown below.


Notice that the red spectrum is that at the maximum position (0,0), and is greatest in amplitude, just as expected. We would expect the spectra colored in orange and green to be more symmetic, as they are on the positive and negative FWHM of the x-direction and so we would expect them to be very similar. Also, in this scan I took only one measurement in the y-direction. This is because the intensity plot showed the y-direction to be more symmetric than the x-direction (see plot from yesterday).

We can see from this plot that there are distinct frequencies at which there is no signal. This may be caused by absorption via water in the air. We might try to reduce this by putting the microscope in an air-tight container and filling it with nitrogen to remove the water.

From this point, I want to look at the relative ratio of signal at each point on the wafer. To do this, I need some sort of a reference spectrum to compare the other spectra to. The way to do this will be to use the (0,0) maximum point as the reference. I want to divide the spectra from each other position into this reference spectrum, and I should look at this in the next few days.

Also, I have added the compilation of the temporal scans, which is shown below. It is not very easy to see each pulse, but this plot is just to show the relative intensity of each position.















So I will hopefully be able to find the ratio of these two signals tomorrow and also take some more data which I can average out to attain more precise spectra.

Wednesday, July 1, 2009

MATLAB and Gaussian Curves

I spent the better part of the day (Wednesday) learning how to use MATLAB. The main source that I used to do this was “MATLAB Guide”, by Desmond J. Higham and Nicholas J. Higham. I made it through the first chapter or so and did some interesting examples including Mandelbrot set, Fibonacci numbers, and simple 2D plots.

I was then able to use MATLAB to input my data from yesterday and plot the two curves on the same set of axes (though it would have been nice to tell MATLAB to put them on orthogonal axes).

Note: I corrected for the x- and y- positions by setting the point of maximum intensity at (0,0).

That data is shown in the image below:









The red shows the data from the x-scan, while the blue corresponds to the y-scan. Notice the final data point at the origin for the y-scan. This is simply because I put the data from the two scans into the same matrix and there were more points for the x-scan than the y, and so the these points were all plotted at (0,0).

Also, below is a plot that Antoine constructed with the same data by using an iterative procedure in MATLAB:








[I apologize for the poor picture quality, but I had to use Paint since I saved these as PDF at lab and Blogger couldn't insert a PDF].

Regardless of the quality, it should be clear that the beam is Gaussian in shape (at least via this method).

I would like to make some more scans to better build up images such as those above, which would probably include scans along y = x and y = -x.

Aside from learning to use MATLAB today, we took a series of scans at specific positions of the excitation beam. We took a scan at maximum (0,0), at maximum x and the lower portion of the HM of y, and then at maximum y with both the upper and lower portions of the HM of x. These were full spectral scans, and each took about 1.5 hours to complete.

Note: There was a major shift in maximum signal change from yesterday to today. The maximum went from about 4.5 mV yesterday to about 7.5 mV today. This may be due to a few things: more power out of the laser OR the excitation beam is locally destroying the Si OR some other thing yet to be determined.

Antoine was going to take one final scan today with the Si wafer slightly shifted so that if the effective change in amplitude was due to a local defect then we should be able to see it right away. For tomorrow, I intend to work some more on learning MATLAB and possibly changing the incident excitation beam on the Si to try and understand the skewness of the x-position (though I think it may have to do with a slight tilt in the wafer (i.e. not perfectly orthogonal to the incident THz beam)). Also, if I have time then I may try to take some more data, as I had mentioned, with the scans being along y = x and y = -x.

Horizontal and Vertical Profiles of THz Beam

I began the day (Tuesday) with trying to center the excitation beam on the Si wafer. I did this by first roughly guessing where the maximum x- and y- displacements were and then performing two manual scans: one along the horizontal with the vertical displacement constant at maximum and then the same thing for the vertical scan (horizontal position constant at about maximum value).

These initial scans I did in increments of 0.5 mm and scanned about 7 mm in the x-direction and about 5 mm in the y-direction. Also, note that I have set x to be horizontal and y to be vertical. The idea of these scans was pretty simple: move the excitation beam across the wafer and record the change in THz signal from the lock-in. The greater the change, the more effective this local mirror was.

Plotting the points from these scans showed some pretty nice looking curves for such quick/crude measurements. Both curves were Gaussian in shape, with some minor anomalies.

Note: These initial measurements were done with an aperture which did two things: limited the amount of THz signal and made it possible to see a Fabry-Perot effect between the aperture and the Si wafer. This is why it was removed in later measurements.

Note: The signal tended to be rather noisy (the signal would oscillate in upwards of 0.1 mV at each position, with the relative maximum being about 4 – 5 mV). To reduce the effective noise, I looked for the signal to ‘oscillate’ and took the maximum and minimum values, later to be averaged together.

I then decided to take more precise measurements of this signal and so I changed the step size to 0.25 mm. I first did this with the x – position and plotted it. I found it to be slightly skewed, but still Gaussian. After this measurement, I removed the aperture and took scans in both the x- and y- directions. I found some very nice looking Gaussian curves for both of these scans, with the x- still being slightly skewed.

These scans ended up being pretty convincing that the THz beam is indeed Gaussian in distribution. I approximated the FWHM of the x- and y- directions to be 2.5 mm and 1.5 mm, respectively. I found the maximum signal difference to be 4.548 mV. The integration constant was 1 s, with 24 dB and 10 mV sensitivity, which were all set on the lock-in.

I also need to check if the excitation beam is stretched too much in the horizontal direction because of its angle of incidence on the wafer. I also need to see if the translation on the stage corresponds to the same translation on the wafer (i.e. if moving the stage 2 mm moves the beam 2 mm).